Content » Vol 74, Issue 3

Investigative Report

TEWL measurement standardization: kinetic and topographic aspects.

Van Sam V, Passet J, Maillols H, Guillot B, Guilhou JJ
DOI: 10.2340/0001555574168170

Abstract

The aim of this study was to standardize the recording technique of the SERVO MED Ep 1C to determine the rest time required for human subjects to obtain stabilized values of transepidermal water loss (TEWL). Measurements of TEWL were performed on the flexor side of the forearm in healthy volunteers. First, different sites were tested on different days to study intra- and inter-individual variations. The recordings were carried out every 5 min for 2 h. Values reached a linear state after 15 min of rest. A time-course type curve of TEWL was established.

Significance

Supplementary content

Comments

Not logged in! You need to login/create an account to comment on articles. Click here to login/create an account.